ABCFarma www.abcfarma.net

Conduction system pacing · Device troubleshooting

Phrenic vs. left bundle capture thresholds in LBBAP: strength-duration physiology and nocturnal diaphragmatic twitching

The requested ratio does not exist in the literature. What does exist is a strength-duration argument that explains when pulse width can help, a hard ceiling that explains when it cannot, and a far more likely explanation for why the twitching happens only at night.

By the Artificial Intelligence Medical Team Published 16 July 2026
The clinical question

In LBBAP patients with nocturnal diaphragmatic twitching, what is the ratio of phrenic nerve capture threshold to left bundle capture threshold at varying pulse widths (0.4–1.5 ms) during supine versus upright posture, and does inspiratory lung volume (diaphragm descent) significantly lower the phrenic threshold, explaining the positional worsening at night?

The honest starting point: this dataset does not exist

No published dataset reports phrenic-to-left-bundle capture threshold ratios across pulse widths, stratified by posture, in left bundle branch area pacing. Producing a strength-duration table that looks authoritative would be a disservice, because the numbers would be invented.

Phrenic nerve stimulation is well characterised in the context of left ventricular and cardiac resynchronisation leads. In LBBAP it is described mainly in case reports and remains under-characterised. The clinically meaningful ratio is lead-specific, position-specific, and patient-specific — it is a measurement, not a lookup.

What follows is what is genuinely established, where the reasoning holds, and where it becomes speculative.

Advertisement

Strength-duration physiology: why pulse width is a lever at all

Both nerve and myocardium obey the same hyperbolic relationship, described by Lapicque:

V(t) = Vrheobase × (1 + chronaxie / t)

The entire manoeuvre rests on one fact: large myelinated nerve has a substantially shorter chronaxie than working myocardium. Roughly 0.1–0.2 ms for phrenic nerve against roughly 0.3–0.7 ms for ventricular myocardium. Skeletal muscle sits between the two.

Geometrically, this means the neural curve flattens early. Beyond approximately 0.4 ms it already sits near its rheobase and gains little from further widening. The myocardial curve is still descending at 1.0–1.5 ms. As the pulse widens, myocardial threshold continues to fall while phrenic threshold has largely bottomed out, and a gap opens between them.

Illustrative strength-duration curves for phrenic nerve and left bundle myocardium Two hyperbolic threshold curves plotted against pulse width from 0.1 to 2.0 milliseconds. The phrenic curve flattens early near 1 volt while the myocardial curve continues to descend, opening a shaded programmable window beyond approximately 0.32 milliseconds. 0 1 2 3 0.5 1.0 1.5 2.0 Pulse width (ms) Threshold (V) curves cross ≈ 0.32 ms Phrenic / diaphragm — chronaxie 0.15 ms Left bundle — chronaxie 0.50 ms Programmable window
Illustrative curves only — not patient data and not derived from any published dataset. Constants chosen for demonstration: myocardial rheobase 0.5 V with chronaxie 0.50 ms; phrenic rheobase 0.9 V with chronaxie 0.15 ms. Note that below approximately 0.32 ms the curves cross and the phrenic threshold sits below the myocardial threshold — the shorter pulse is the worse choice.
Illustrative model output across the pulse widths in question
Pulse widthMyocardiumPhrenicRatioWindow
0.4 ms1.13 V1.24 V1.10×0.11 V
0.8 ms0.81 V1.07 V1.32×0.26 V
1.0 ms0.75 V1.04 V1.38×0.29 V
1.5 ms0.67 V0.99 V1.48×0.32 V

The rheobase ceiling: the cautionary half of the argument

The ratio at any pulse width is:

R(t) = (Vrheo,n / Vrheo,m) × (1 + cn/t) / (1 + cm/t)

As pulse width tends to infinity, the second term tends to 1 and the ratio converges on the ratio of the two rheobases. This yields three consequences that matter more than any tabulated number.

The ceiling

Rheobase ratio sets the ceiling; chronaxie only determines how quickly that ceiling is approached. Widening the pulse can approach the asymptote but can never exceed it.

Two caveats to the model

  1. It assumes neural capture. If the culprit is direct diaphragmatic muscle capture rather than phrenic nerve capture, the chronaxie gap narrows considerably — skeletal muscle sits far closer to myocardium — and the pulse-width manoeuvre largely evaporates.
  2. The real curves are not stationary. Posture and respiration displace the phrenic curve vertically by changing distance. A window measured supine at rest may close during a particular phase of the respiratory cycle. This is precisely why a bench-derived ratio would not help even if one existed.
Advertisement

Posture and the nocturnal pattern

The postural component of the question is well grounded and mirrors what is documented for left ventricular and resynchronisation leads. Postural phrenic capture is a recognised phenomenon: diaphragmatic stimulation appears on lying down that is absent when upright.

Supine and lateral positions alter the geometry between lead, phrenic nerve, and diaphragm. When supine, abdominal contents displace the diaphragm cephalad, functional residual capacity falls, and the diaphragm sits higher — closer to a septal pacing source.

A correction on direction

The inspiratory hypothesis is plausibly reversed. During inspiration the diaphragm descends — moving away from a septal source, which would generally raise the threshold for direct diaphragmatic capture rather than lower it. The nocturnal worsening is better explained by the tonic supine change than by inspiratory descent.

Respiration does modulate the threshold dynamically: capture can wax and wane across the respiratory cycle as anatomy shifts millimetre by millimetre. Whether the net inspiratory effect raises or lowers threshold depends on whether the culprit is true phrenic nerve capture, which is relatively fixed anatomically, or direct diaphragmatic capture, which is far more geometry-dependent.

An anatomical objection worth raising first

A mid-septal LBBAP lead is an unlikely place from which to capture the phrenic nerve. The right phrenic nerve descends along the superior vena cava and lateral right atrium. The left phrenic courses over the left ventricular lateral wall and pericardium. Neither approaches the mid-interventricular septum.

Where diaphragmatic twitching genuinely arises from an LBBAP lead, the likelier explanations, in rough order, are:

A programming decision tree

Step 1 — Establish that the device is responsible

Step 2 — Localise the lead and the electrode

Step 3 — Programming levers, in order of yield

  1. Reduce output. Devices are commonly programmed with generous headroom. A bundle threshold of 0.6 V paced at 2.5 V leaves considerable room to move.
  2. Change the vector. The highest-yield single manoeuvre where anodal capture is involved.
  3. Only then exploit the strength-duration curve. Measure the actual threshold pair at two or three pulse widths, in the posture that provokes the symptom, and determine whether a window exists. This is done in the patient, never from a table.

Step 4 — The step most often skipped

Highest-yield question for a nocturnal presentation

Establish whether ventricular pacing is required at night at all.

Where atrioventricular conduction is intact, the elegant answer is not to pace:

This hypothesis should be tested before any investment in strength-duration work. Review the percentage of ventricular pacing by time of day in the device diagnostics. If ventricular pacing is minimal during waking hours and near-continuous between midnight and 06:00, the answer lies in lower rate limit and AV delay programming rather than in capture physics.

Step 5 — Where programming fails

Lead revision. Where the rheobase relationship is unfavourable — the crossed-curves scenario — no amount of programming will separate the two thresholds, and continued attempts merely postpone the inevitable while sleep remains disrupted.

Summary

Educational content. This material is intended for clinicians and biomedical professionals and is provided for education only. It does not constitute medical advice and must not substitute for individualised clinical judgement, device manufacturer specifications, or institutional protocol. The strength-duration constants used here are illustrative and are not derived from any patient dataset or published series; they demonstrate the shape of the relationship, not its magnitude in any individual. Device programming decisions must be made by qualified personnel with access to the patient and to full device diagnostics.
Advertisement