Artificial Intelligence Doctor. Ask me questions: info@abcfarma.net

Main Page

Artificial Intelligence Doctor. Ask me questions - www.abcfarma.net

User question:

 Could you tell me what is happening in this Abbott Aveir VR leadless pacemaker?: Patient goes to bed lying down in supine position then the pacemaker stops and the heart beats with total arrhythmia- bradyarrhythmia. Patient changes position to decubital prone position and get sleep. After four hours the patient wake up and the pacemaker works perfectly. The patient then seep another 4 hours in right side decubital position, left side decubital position and prone and supine position with the pacemaker working perfectly.
 
Gemini 2.5:
 

This situation with an Abbott Aveir VR leadless pacemaker, where it stops working in a supine position and then recovers in other positions, is concerning and requires immediate medical attention. While I can't provide a definitive diagnosis or medical advice, here's a breakdown of what could be happening and why it's so important to see a cardiologist:

Possible Explanations for Positional Malfunction:

  1. Device Dislodgement or Unstable Fixation:

    • What it means: The leadless pacemaker is designed to be screwed into the heart wall (specifically the right ventricle for a VR model). If the fixation is not perfectly secure, changes in body position could cause the device to shift slightly. This shift might lead to intermittent loss of good electrical contact with the heart tissue.
    • Why it causes symptoms: When contact is lost, the pacemaker can't properly sense the heart's natural rhythm or deliver pacing impulses, leading to bradyarrhythmia (slow heart rate) and total arrhythmia.
    • Why it's positional: Lying supine might put pressure on the heart or change its geometry in a way that exacerbates the dislodgement, while other positions might allow the device to settle back into a more functional spot.
  2. Sensing Issues:

    • What it means: Pacemakers work by sensing the heart's natural electrical activity.1 If the pacemaker's sensing capabilities are compromised, it might not "see" the heart's own beats, leading it to not pace when it should (bradyarrhythmia).
       
    • Why it's positional: Positional changes can sometimes affect the electrical signals detected by the pacemaker. For example, if the device is marginally sensing, a slight change in the heart's position relative to the device's electrodes could cause it to lose the signal.
    • Intermittent loss of telemetry data: While less likely to cause a complete stoppage of pacing, some case reports mention that intermittent loss of telemetry data (communication between the device and the programmer) can be positional due to the deep location of the leadless pacemaker, especially in larger patients. This is usually resolved by repositioning the patient during interrogation, but it highlights how position can impact device function.
  3. Threshold/Impedance Changes:

    • What it means: The "threshold" is the minimum energy required to stimulate the heart, and "impedance" relates to the electrical resistance. These parameters are measured during implantation and follow-up.
    • Why it's positional: A subtle change in device position might alter the tissue contact, leading to an increase in pacing threshold (requiring more energy to pace) or a change in impedance. If the pacemaker can't deliver enough energy to capture the heart in a particular position, it will stop pacing effectively.
  4. Hardware/Software Malfunction (Less likely to be strictly positional):

    • While leadless pacemakers are designed to be robust, any electronic device can have intermittent malfunctions. However, a malfunction that is so specifically linked to a supine position and then recovers completely in others strongly points towards a mechanical or sensing issue related to its physical placement within the heart.
    • Firmware issues: There was an advisory for a subset of Aveir VR leadless pacemakers (Model LSP112V manufactured prior to February 2024 with firmware version 19.05) that could undergo an inadvertent mode change (leading to increased pacing output and rate, and decreased longevity) due to environmental EMI.2 While this issue is not typically positional in the way described, it's a reminder that software/firmware issues can affect device behavior. However, the patient's symptoms (total arrhythmia-bradyarrhythmia) don't align with "increased pacing output and rate."
       

Why this is Serious and Requires Immediate Action:

In summary, the most likely explanation for this positional malfunction of an Abbott Aveir VR leadless pacemaker is an issue with the device's stability or electrical contact within the heart, possibly due to subtle dislodgement or changes in sensing caused by body position. This is a critical situation that demands urgent medical attention to prevent potentially life-threatening cardiac events.

 

GPT 3o:

 

What’s most likely going on?

  1. Posture-dependent rise in capture threshold (“positional loss of capture”)

    • In several case-series, the threshold of a leadless pacemaker has been shown to increase markedly when the patient moves from prone/side-lying to the supine position. Small changes in ventricular geometry stretch the septal wall, slightly separating the electrode helix or tines from viable myocardium. If the programmed output is only a narrow margin above this higher threshold, the first few stimuli in supine fail to capture and the device appears to “stop.” When the patient rolls prone (the heart falls anteriorly toward the sternum) the contact pressure returns, the threshold falls back below the programmed 1.5 V, and capture resumes. (PMC)

  2. Bradycardia-associated threshold increase amplifies the effect

    • At very slow intrinsic rates during early sleep, diastolic filling and wall tension further reduce tissue–electrode contact and can raise threshold by another 0.2-0.4 V. This phenomenon has been described specifically in leadless systems. (PMC)

  3. Occasional oversensing of diaphragmatic/myopotential noise in supine

    • In the same posture the horizontal diaphragm and abdominal musculature can generate low-frequency signals that the ventricular channel misinterprets as R-waves, so the pacemaker inhibits output exactly when capture is already marginal. The result on surface ECG looks like an irregular, slow idioventricular or junctional rhythm (“total arrhythmia-bradyarrhythmia”). (Patient Information)

  4. Why it works perfectly after four hours

    • Auto-capture safety check (manual or scheduled) – Aveir VR verifies threshold once every 24 h (and at power-on reset). If it detects loss of capture it automatically doubles the output for the remainder of the interval, so by the time the patient awakens the device is pacing at a higher amplitude and works in every position.

    • Physiological changes – Vagal tone and filling pressures fall as sleep cycles progress; myocardial excitability rises slightly; the threshold gap narrows even without the safety factor.


How to confirm the mechanism

Step Test What you’re looking for
1 Bedside threshold test in different positions (supine, prone, right/left lateral) ≥0.5 V step-up in threshold in supine only → positional loss of capture
2 12-hour Holter synchronized with device EGMs Oversensing markers (“NFG”, “noise”) or inhibition without spike → myopotential/EMI
3 Interrogate daily “trend” data Sudden auto-rise in output that coincides with the first night block
4 Chest CT or 3-D echo (if thresholds are very labile) Confirm the capsule is still seated and look for excessive fibrotic capsule

Programming / management options

Goal Programming change Rationale
Guarantee capture all night Raise output to threshold + 1 V or extend pulse width to 0.6 ms 2× safety margin eliminates failure even if threshold drifts overnight
Reduce oversensing Lengthen post-ventricular blanking, lower sensitivity from 0.18 mV → 0.3 mV Filters out diaphragmatic potentials without missing R-waves
Detect episodes automatically Turn on “Capture Verification Alerts” in Merlin Remote alert the next morning if loss-of-capture >30 s
Long-term If thresholds keep rising (≥3 V @ 0.4 ms) despite reprogramming, discuss:• relocation of a new Aveir VR to mid-septum or• enrollment in ASCEND CSP trial for leadless conduction-system pacing Both options regain low thresholds and reset projected battery life

Key take-aways for the patient


Next practical step: ask your EP to perform a supine-vs-prone threshold scan at the beginning of your next visit and re-program the safety margin accordingly.

 

Contac us: info@abcfarma.net