Abbott Aveir VR Leadless Pacemaker Intermittent Capture Issue

Artificial Intelligence Doctor. Ask me questions: info@abcfarma.net

Main Page

Artificial Intelligence Doctor. Ask me questions - www.abcfarma.net

User question:

Prompt for Analysis: Abbott Aveir VR Leadless Pacemaker Intermittent Capture Issue

Device Details:

  • Model: Abbott Aveir VR (leadless pacemaker)

  • Current Settings:

    • Mode: VVI

    • Base Rate: 50 bpm

    • Max Sensor Rate: 130 bpm (Passive sensor)

    • Capture Threshold: 3.0 V @ 0.4 ms

    • Sensing Threshold: 4.0 mV

    • Impedance: 550 Ohms

Patient Symptoms:

  • Some nights, the device fails to capture, resulting in the patient only sleeping 3 hours (suggesting possible loss of pacing).

  • Other nights, the patient sleeps well for 10 hours straight (suggesting normal device function).

Questions for Analysis:

  1. Why might the Aveir VR intermittently lose capture?

    • Could the 3.0 V @ 0.4 ms threshold be borderline due to changes in myocardial contact or micro-dislodgement?

    • Is there evidence of threshold rise at night (e.g., due to sleep apnea, positional changes, or metabolic factors)?

    • Could high impedance (550 Ω) suggest poor contact or fibrosis at the electrode-tissue interface?

  2. Could undersensing contribute to symptoms?

    • The sensing threshold is set at 4.0 mV—is this sufficient, or could intrinsic signals be missed, leading to inappropriate pacing inhibition?

  3. Does the VVI mode play a role?

    • Since the device is VVI with a passive sensor, could the lack of rate responsiveness contribute to symptoms (e.g., nocturnal bradycardia)?

    • Should a rate-responsive mode (VVIR) be considered to improve sleep-related hemodynamics?

  4. What diagnostics should be checked?

    • Capture threshold trends (night vs. day variations).

    • Event markers/EGMs from nights with poor sleep to confirm loss of capture.

    • Impedance trends (stability vs. fluctuations).

  5. Possible Interventions:

    • Increase output safety margin (e.g., 3.5 V @ 0.4 ms or adjust pulse width).

    • Re-evaluate sensing settings if undersensing is suspected.

    • Assess for sleep apnea or positional micro-leads movement (e.g., via Holter or sleep study).

    • Consider reprogramming to VVIR if chronotropic incompetence is suspected.

Final Request:
Based on the device settings and symptoms, what is the most likely cause of intermittent loss of capture, and what would be the recommended troubleshooting steps?


This prompt ensures a structured analysis of the device behavior, potential causes, and next steps for clinical evaluation. Let me know if you'd like any refinements!

 

what is the most likely cause of intermittent loss of capture, and what would be the recommended troubleshooting steps?
 

 

Most Likely Cause of Intermittent Loss of Capture in the Abbott Aveir VR Pacemaker

Given the symptoms (intermittent failure to capture, leading to poor sleep some nights) and the current settings (3.0 V @ 0.4 ms, impedance 550 Ω), the most likely causes are:

  1. Borderline Capture Threshold with Nocturnal Threshold Rise

    • The 3.0 V @ 0.4 ms setting may be too close to the actual threshold, especially at night when factors like:

      • Sleep apnea (hypoxia increases pacing threshold).

      • Positional changes (e.g., lying on the left side causing micro-dislodgement).

      • Metabolic variations (e.g., electrolyte shifts, vagal tone changes).

    • If the threshold rises at night, the pacemaker may fail to capture intermittently, explaining why some nights are worse than others.

  2. Possible Micro-Dislodgement or Fibrosis at Lead-Tissue Interface

    • The impedance of 550 Ω is on the higher side (typical range for leadless pacemakers is ~400–800 Ω, but trends matter).

    • If impedance is increasing over time, it could suggest fibrosis or poor contact, leading to intermittent capture failure.

  3. Undersensing Leading to Missed Pacing (Less Likely, but Possible)

    • The sensing threshold of 4.0 mV is reasonable, but if intrinsic R-waves vary at night (e.g., due to PVCs or low-amplitude signals), undersensing could inhibit pacing when needed.


Recommended Troubleshooting Steps

1. Immediate Device Interrogation & Diagnostics

  • Review nightly auto-capture threshold trends (if available in diagnostics).

  • Check stored EGMs/event markers from nights with poor sleep to confirm loss of capture.

  • Look for impedance trends (sudden changes suggest micro-dislodgement).

2. Adjust Pacing Output for Safety Margin

  • Increase output to 3.5 V @ 0.4 ms (or 3.0 V @ 0.6 ms) to ensure adequate safety margin.

  • If threshold is borderline, Abbott’s AutoCapture™ algorithm (if enabled) may help, but manual adjustment is safer if intermittent failure is confirmed.

3. Evaluate for Sleep Apnea or Positional Changes

  • If the patient has sleep apnea, nocturnal hypoxia could increase pacing thresholds—consider a sleep study.

  • If symptoms are position-dependent, assess whether certain sleeping positions affect capture.

4. Consider Reprogramming to VVIR (If Chronotropic Incompetence is Suspected)

  • Since the sensor is passive, the pacemaker does not adjust rate for metabolic demand.

  • If the patient has poor heart rate response at night, VVIR mode might help maintain better hemodynamics.

5. If Problem Persists: Re-evaluate Lead Position

  • If impedance trends suggest micro-dislodgement or fibrosis, a fluoroscopy check may be needed to confirm lead stability.

  • In rare cases, repositioning or replacement may be required.


Final Recommendation

  • First step: Increase pacing output (e.g., 3.5 V @ 0.4 ms) and monitor symptoms.

  • Second step: Check diagnostics for nocturnal threshold rise or impedance changes.

  • Third step: If no improvement, evaluate for sleep apnea or consider VVIR mode.

Would you like any additional details on specific diagnostics or Abbott’s AutoCapture algorithm?

 
 

Contac us: info@abcfarma.net