Devices with automatic threshold tracking
| Device family | Feature name | What it does |
|---|---|---|
| Micra VR (MC1VR01) | Ventricular Capture Management (VCM) / Capture Management | Automatically measures RV capture threshold and adjusts output with a programmable safety margin; periodic checks support energy savings and consistent capture. |
| Micra AV (MC1AVR1) | Capture Management (RV) | As above, plus atrial mechanical sensing for AV synchrony (not part of threshold tracking). |
| Micra VR2 / Micra AV2 | Capture Management (RV); enhanced algorithms in AV2 for atrial mechanical sensing | Successors with extended longevity; maintain automatic RV threshold management functionality. |
Notes on AVEIR™ devices
What is documented: AVEIR VR/DR materials highlight pre‑fixation mapping (commanded EGM, impedance, and initial capture threshold) to choose the implant site and support retrieval/re‑positioning. Routine follow‑up includes programmer‑based checks.
What is not clearly documented: a continuously running, implanted automatic threshold‑tracking algorithm (like VCM) that autonomously updates output post‑implant. If present in future software, consult the latest manuals.
Why this matters for nocturnal variation
- Automatic threshold tracking can detect night‑time rises and maintain capture with small, time‑limited output increases.
- Adaptive margins help optimize longevity by lowering output when thresholds are reduced during daytime.
- Program within each model’s approved options; verify logs for timing/frequency of threshold checks.
References (selected)
- Medtronic Academy: Ventricular Capture Management (VCM) overview.
- FDA SSED for Micra (P150033) detailing VCM objective.
- Micra AV Implant Manual noting Capture Management.
- Abbott AVEIR VR/DR pages and SSED describing mapping and implant measurements.